Electronic Test Instruments - Signal / Sweep Generator
S-900 AM/FM Signal Generator


- 100KHz - 55MHz
useable upto 110MHz
(2nd Harmonic)
 
- Frequency Counter, 55MHz
- TTL Output
- Low radiation


S-900 AM/FM Signal Generator
is a low cost equipment ideally suited for Education and Service applications, it comprises of the following.

AM/FM Signal Generator, which generates frequencies from 100KHz to 55MHz in 6 bands. The second harmonic of the last band generates frequencies upto 110MHz.

Frequency Counter which measures frequency of the AM/FM and External input.

TTL Output is available at the frequency readout on the counter. TTL output can be used to drive various circuits during experimentation.

Characteristics

AM OSCILLATOR

Frequency Range:  
1) 100KHz - 300KHz  
2) 300KHz - 1MHz  
3) 1MHz - 3MHz  
4) 3MHz - 10MHz  
5) 10MHz - 25MHz  
6) 25MHz - 55MHz  
continuously variable over the whole range with a coarse and fine variable
   
Modulation Frequency 400Hz approx
Modulation depth a) Internal Modulation 30% (electronically adjusted
  b) External Modulation - adjustable.
Output Amplitude 100mV p-p +2dB into 75 ohms, corresponding to a carrier voltage of 27mV rms approx
   
FM OSCILLATOR  
3MHz - 10MHz  
10MHz - 25MHz  
25MHz - 55MHz  
80MHz - 110MHz (2nd harmonic of band 6)
continuously variable over the whole range with a coarse and fine variable
Modulation Frequency 400 Hz approx
Frequency Deviation a) Internal - adjustable 0-22.5KHz from 5MHz to 55MHz
  b) External - adjustable 0-40KHz from 3MHz to 55MHz for Audio input frequency range 300Hz to 5KHz and a max input of 8V p-p
Output Amplitude 75mV p-p +2dB into 75 ohms
Attenuation - a) Step Attenuator 5 steps (20dB) each
  b) Variable attenuator
   
FREQUENCY COUNTER
Readout 5 digit LED display
Range Selection MHz/KHz/Hz
Time Base 10MHz Xtal derived
Source Selection a) Internal AM
  b) Internal FM
  c) External frequency
External Input 10Hz to 55MHz
Sensitivity 20mV rms at 1KHz
   
TTL OUTPUT
Output Voltage TTL Level approx when terminated in 75ohms, abberations may exceed 40%
Frequency Range 100KHz to 55MHz
   
GENERAL  
Mains Voltage 230V +10%, 50Hz
Dimensions 378 (W),144 (H),190 (D)
Accesories Operating Manual, Cable
 

Remarks: All accuracies specified at 25deg C ambient, 230V mains, 50Hz with instrument grounded - after warmup of 30 min. Specifications, illustrations and price change privileges reserved.